• HIGH: High level limit, in seconds, that the peak crest can shift LIMITS: Concentration (Reported Component Values)
• LOW-LOW: Sets the low-low indicator concentration limit
• LOW: Sets the low indicator concentration limit
• HIGH: Sets the high indicator concentration limit
• HIGH-HIGH: Sets the high-high indicator concentration limit
The Status subtab displays the component’s associated indicators, their state and allows reset. The
component indicators must be checked active on the Status Tab to see state color.
4.6.5 Streams
Select Streams in the Function Select list to display all streams configured for the analyzer (see
Figure 4-28). Streams can be created, configured and deleted from this display.
Figure 4-28: Setup>Streams Configuration
STREAM LIST: Drop-down list of available streams
STREAM NAME: Reference name of the stream
STREAM ONLINE: Current status of Stream (Checked = Active/Online). Clearing (Deactivating)
triggers the Offline Indicator.
STREAM VARIANT: Current variant of the stream being used
RUNTIME TYPE: Drop-down list of selectable stream types
VALVE NAME: Drop-down list of selectable valve types
VALVE ADDRESS: Reference address of the selected valve
The Status subtab displays the configured indicator’s state, date and time of the selected stream.
To create a stream:
1. Navigate to the Setup tab>Config subtab and select Streams in the Function Select list.
2. Select the Add New icon and enter a name and a variant for the new stream. Stream names
must be unique in the analyzer.
3. Confirm the stream creation by pressing the Accept button, or cancel the operation by pressing
the Abort button.
4. Select the RUNTIME TYPE from the drop-down list.
5. Select the VALVE NAME from the drop-down list.
6. Select the STREAM ONLINE box to activate the stream in the schedule.
To delete a stream:
1. Navigate to the Setup Tab>Config subtab and select Streams in the Function Select list.
2. Select the stream to delete from the STREAM LIST drop-down list.
3. Select the Delete icon.
4. Confirm the stream deletion by pressing the Accept button, or cancel the operation by pressing
the Abort button.
4.6.6 Master controller
Select Master Controller in the Function Select list to display the Config and Status subtabs and all
associated cards and peripherals (see Figure 4-29).

Ovens with Integrated Controllers will show a Master Controller in the Function Select list, even
though no physical PGC5000A Master Controller is present. The SBC subtab will be under the Master
Controller button. The Power Supply, Purge Control, and Front Panel subtabs will not be shown, as
these Master Controller parameters are not needed in the oven with Integrated Controller.
The associated items in the Function Select list are shown in an indented list. If an associated item or
device contains or has a subset of parts or cards, those will be shown in an indented button list below
that item or device.
Master Controller cards: As with the Master Controller button, each associated card in the list
has two subtabs: Config and Status. The Config subtab lists relevant or configurable
information about the card, while the Status subtab is blank on these items. The table below
provides more information.
Table 4-3: Master controller components
Button/Peripheral Config (Displays)
Power Supply Part number
Front Panel Part number
Purge Control Part number
Singe Board Computer (SBC) Serial and part number; Current version software
Network Adapters Network Address Settings: See Network Setup Figure 4-30 in this
section
USB Hub User editable port identification for reference purposes
Clock System Time and Date Entry: See Clock Setup Figure 4-32 this
section
Malfunction Alarm Allows user-editable port identification for reference purposes
Purge Alarm Contact type and rating. Editable field for reference purposes
Config: This subtab displays the configuration of the selected item.
Status: This subtab displays the state of configured indicators. If triggered (state changed), it
displays that event date and time.
IMPORTANT NOTE: Do not check the SECURED box unless the ACL contains at least one
Supervisor user.
When the SECURED box is selected, no changes can be made to the analyzer’s functions unless the
user is logged in with Supervisor privileges. Once the analyzer is secured, the SECURED box can only
be cleared by a supervisor.