+086-15305925923

K-WANG

Service expert in industrial control field!
NameDescriptionContent
Adequate Inventory, Timely Service
pursuit of excellence 
Ship control system
Equipment control system
Power monitoring system
Current position:
新闻动态
newS
   
Brand

TEKTRONIX TCPA300/400 current probe amplifier

From: | Author:Wang | Time :2025-11-21 | 343 visit: | 🔊 Click to read aloud ❚❚ | Share:

TEKTRONIX TCPA300/400 current probe amplifier

Product positioning

Product portfolio:

Amplifier: TCPA300 (supports TCP312/305/303 probes), TCPA400 (only supports TCP404XL probes);

Probe: TCP312(30A/100MHz)、TCP305(50A/50MHz)、TCP303(150A/15MHz)、TCP404XL(750A/2MHz, Discontinuous).


Safe operation and system configuration

1. Core security standards

Prohibited live operation: Before connecting/disconnecting the probe/test line, the voltage source must be cut off to avoid insulation breakdown inside the probe;

Terminal matching requirements: The amplifier output needs to be connected to a 50 Ω load (oscilloscope or 50 Ω through terminal), otherwise the NOT TERMINATED IN 50 Ω LED will light up, affecting measurement accuracy;

Wire usage restrictions:

TCP312/305: Only insulated wires can be tested, and contact with exposed conductors is prohibited;

TCP303/404XL: can measure exposed conductors, but the probe needs to be removed and installed after power failure;

Environmental restrictions: Prohibit humid/explosive environments, working humidity of 5% -95% (≤ 30 ℃), 5% -85% (>30 ℃ -+50 ℃), non condensing.

2. System configuration and connection

(1) System composition

A complete current measurement system requires three parts:

TCPA300/400 amplifier: converts probe induced current into proportional voltage;

TCP300/400 series probe: clamp type induction current (effective after sliding cover locking);

Oscilloscope: requires a 50 Ω input impedance (if not available, add a 50 Ω straight through terminal), and the recommended bandwidth is 5 times the probe (e.g. TCP312 requires ≥ 500MHz oscilloscope).

(2) Connection steps

Amplifier → Oscilloscope:

TEKPROBE compatible oscilloscope: use TEKPROBE interface cable;

Non compatible oscilloscope: use 50 Ω BNC wire, set the oscilloscope input to 50 Ω;

Probe → Amplifier: Align the red marked points of the probe and amplifier, and lock them in place (do not twist);

Preheating: All equipment should be powered on and preheated for 20 minutes to ensure thermal stability.

Core Function Operation Guide

1. Key operations of the probe

(1) Demagnetization and automatic balancing (mandatory steps)

Function: Eliminate residual magnetism in the probe core, offset DC offset in the amplifier, and avoid measurement errors;

Trigger conditions: The probe slide is locked and disconnected from the live conductor;

Operation steps:

Press the PROBE DEGAUSS AUTOBALANCE button on the amplifier;

Wait for 5 seconds, the LED will flash from orange to green (successful), red (failed, terminal/probe connection needs to be checked);

Trigger scenarios: after startup, replacement of probe, overload, exposure to strong magnetic field.

(2) Sliding operation (different probe differences)

Probe model, sliding cover, opening method, locking method, precautions

TCP312/305 pulls the sliding cover backwards and pushes it forward until the buckle is fixed, only clamping the insulated wire

TCP303/404XL press lock+press handle release handle+press lock can clamp exposed wires (power-off operation)

2. Amplifier core control

(1) Key buttons and indicators

Control/indication function description exception handling

The COUPLING key switches between AC/DC coupling (AC only measures AC, DC measures AC/DC), and flashing indicates the display of Error Code

The RANGE key (TCPA300) switches the range (1/5/10/50A/V), and the LED indicates the current range. If the LED is not on, it means that the probe is not connected

PROBE OPEN LED probe slide not locked and turns off after locking the slide

OVERLOAD LED red: current beyond continuous range; Orange: Overtemperature; Flashing: Both exceed power-off cooling+demagnetization

NOT TERMINATED LED amplifier output not connected to 50 Ω load. Check oscilloscope impedance or add terminal

(2) Manual balancing (under DC coupling)

Function: Fine tune DC offset to improve measurement accuracy;

Operation: Press the MANUAL BALANCE key (left and right keys to adjust the offset direction), the LED lights up to indicate entering manual mode;

Limitation: Only DC coupling is effective, disabled under AC coupling.

3. Measurement process (taking DC current as an example)

Preheat the system for 20 minutes, set the amplifier to DC coupling, and set the oscilloscope input to 50 Ω;

Probe slide lock (no conductor), perform demagnetization/automatic balancing;

Adjust the grounding reference of the oscilloscope to the target scale;

Open the probe slide cover, clamp the tested conductor (probe arrow pointing in the normal current direction: positive → negative), and lock the slide cover;

Adjust the oscilloscope time base, trigger and gain, and read the current value (current=oscilloscope voltage value/amplifier range, for example, 2V corresponds to 10A at 5A/V range).

Performance validation and calibration

1. Preconditions for Performance Verification

Environmental conditions: temperature 23 ℃± 5 ℃, humidity meets requirements, no strong magnetic field;

Equipment requirements: current source (0.1% accuracy), DMM (5.5bit resolution), high-frequency current loop, etc. (see Table 5-3/5-11/5-17 for details);

Data recording: Corresponding test record forms (such as TCPA300 Table 5-16, TCP305 Table 5-29) need to be filled out.

2. Core testing items (amplifier+probe)

(1) DC gain accuracy test (taking TCP305 as an example)

Equipment connection: current source → 5-turn current loop → probe → amplifier → DMM (50 Ω terminal);

Test steps:

After demagnetization, the probe clamps the current loop and the current source outputs+1.00A;

Record the DMM voltage value M1, and then output -1.00A to record M2;

Calculation error:% Error=[(M1-M2) -2 × VE]/(2 × VE) × 100% (VE is the expected voltage, such as 1A corresponding to 1V in the 5A/V range);

Qualified range: ≤± 3% (guaranteed value), ≤± 1% (typical value).

(2) Bandwidth test (using TCP312 as an example)

Equipment connection: level sine wave generator → high-frequency current loop → probe → amplifier → oscilloscope;

Test steps:

The generator outputs a 3MHz, 3Vpp signal and records the oscilloscope peak to peak value M1;

Adjust the generator frequency to 100MHz (TCP312 bandwidth) and record M2;

Verification: (M2/M1) × 1.18 ≥ 0.707 (1.18 is the impedance correction factor);

Qualified range: -3dB bandwidth ≥ 100MHz.

(3) Rise time test (taking TCP404XL as an example)

Equipment connection: High voltage pulse generator → Built in current loop → Probe → Amplifier → Oscilloscope;

Test steps:

The generator outputs 5A, 250 μ s pulses, and the oscilloscope is set to 80ns/div;

Measure the rise time of pulses ranging from 10% to 90%;

Qualified range: ≤ 175ns (derived from a bandwidth of 2MHz: 0.35/2MHz=175ns).

3. Calibration adjustment (only operated by qualified personnel)

(1) Amplifier adjustment (TCPA300)

Prerequisite: The casing needs to be disassembled and a calibration adapter (174-4765-00) needs to be used;

Step: The current source outputs 0.100A, adjust the corresponding range potentiometer (such as adjusting the 25 Ω potentiometer for 5A/V) to make the DMM voltage meet the expected value (such as 0.1A corresponding to 0.25V at 5A/V range).

(2) Probe adjustment (TCP305/312)

DC gain adjustment: Open the probe housing and adjust the DC gain potentiometer to make the DMM voltage error ≤ ± 3% (refer to Table 6-7);

Attention: Record the original position before adjustment to avoid exceeding the specifications.


Troubleshooting and Maintenance

1. Common faults and solutions

Possible causes and solutions for the fault phenomenon

Demagnetization failed (LED red), 50 Ω load not connected, probe not locked, check terminal+lock slide cover

OVERLOAD LED constantly on, current exceeding range, probe overheating, current reduction+power-off cooling+demagnetization

The measurement value is not accurate without demagnetization, the range is selected incorrectly, the terminal is not connected to demagnetization+check the range+check the 50 Ω load

The probe without output signal is not connected, and the coupling is set to AC (measuring DC). Connect the probe and switch the DC coupling

2. Interpretation of Error Code

Display mode: AC/DC coupled LED flashes alternately, and 4 fault LEDs (PROBE OPEN is MSB, NONCOMPATIBLE is LSB) display a 4-digit binary code;

Common code:

Code 1: No Hall element detected in the probe → Reconnect or replace the probe;

Code 2: DC offset calibration failed → restart amplifier+demagnetization;

Code 15: Internal software error - If restarting is ineffective, contact after-sales service.

3. Daily maintenance

Cleaning: Wipe the outer shell with 75% isopropanol, and do not use solvents such as benzene/acetone;

Storage: Non working temperature -40 ℃ -+75 ℃, avoid moisture/dust;

Calibration cycle: It is recommended to calibrate once a year, with a reference frequency error (± 10ppm) valid for one year.


Accessories and Replacement Parts

1. Standard accessories (included with the product)

Category, accessory name, purpose

Amplifier accessory 50 Ω straight through terminal (011-0049-02) matching output impedance

Connect the TEKPROBE interface cable (012-1605-00) to the TEKPROBE oscilloscope

BNC line (012-0117-00) connected to non TEKPROBE oscilloscope

Probe accessories Probe cover (016-1923-00) protects the probe

Grounding lead (196-3120-01) probe grounding

2. Key replacement parts (partial)

Component type, model, and part number

TCPA300 motherboard circuit motherboard 671-5434-00

TCP305 coil current transformer component 120-184-00

TCP404XL cable probe cable assembly 174-4816-00

Amplifier fan axial fan (12VDC) 119-6721-00


  • Metso XZB10025-011T12 - Moog Valve 9650962700
  • Metso XZB10025-010T12 - Moog Valve 9650965300
  • Metso-SP Transmitter Cable - Operating Unit with 10m Cable
  • Harkila Metso Hybrid - Hunting Jacket Willow Green
  • Metso ND8221/S1 - Neles Valve Positioner
  • Metso METSO10119 - ScreenCheck Component
  • Metso LK-2025 - Linkage Kit
  • Metso A416733 - IQweight External Air Gap Temperature Sensor
  • Metso RKC22MT - Automation Component
  • Metso A413281 - Automation Board
  • Metso LCP9HLY - Stainless Steel Enclosure
  • Metso NP704S/B1S1 - Pneumatic Positioner with I/P Converter
  • Metso BAH3 - Slurry Seal Type
  • Metso A413531-03 - PLC Module Chassis Rack
  • Metso P4610004 - MCAI Communicator Controller
  • Metso QN2AN03HDM - Quartz Valve Positioner Stonel
  • Metso RKW-352-XZ - Automation Component
  • Metso A4300081 - Kajaani LCD Display
  • Metso A428026 7B - Kajaani Rectifier Unit 1
  • Metso 202218 - Kajaani Sensor Power Supply SR 902660
  • Metso 9150 316SS CSTL - Jamesbury 150-2" BN50 Valve
  • Metso 81806 - IOP101 Module
  • Metso STOM000272 - Paper Bushing CR38
  • Metso 1975413 - Chain Hoist Links
  • Metso IMO G053-191597 - Minor Seal Kit
  • Metso VAL0087418 - Cylinder Valmet NTS6LV
  • Metso A413061 - DMU Board VPA 420355-4B
  • Metso P2620-80/25-100 - Automation Component
  • Metso VAL0035292 - Automation Component
  • Metso A413721 - Automation Component
  • Metso ST600 - Automation Component
  • Metso P4500001 - Automation Component
  • Metso SDBX/RA2 - CAT x KVM Extender
  • Metso D201126-DI8 - Digital Input Module
  • Metso D201189L - AI8H Analog Module
  • Metso NA004418 LK1016 - Wafersphere Actuator Mounting Bracket
  • Metso RKN172MTT - Repair Kit
  • Metso MM0311584 - Air Filter Kit RadialSeal
  • Metso 181517 - IOP301 Isolated Analog Input Module
  • Metso RAU3906506 - Automation Component
  • Metso 181208 - IOP332 Digital Input Module
  • Metso A413760 - AIF2C PCB Module
  • Metso 006 1089 36 - Shaft Bearing Seal
  • Metso A413046 - Automation Component
  • Metso VKJ650 - Automation Brake Component
  • Metso LK-1875 - Linkage Kit
  • Metso D201193L - FI4S24 Frequency Input Module
  • Metso D200533 - BIU82 Binary Input Module
  • Metso 500838-M1 - Impeller Pump Release Collar Assembly
  • Metso 2.00SB2ALUS19A - Pneumatic Cylinder 1.375 Stroke
  • Metso D201351 - RES Reserve Module
  • Metso A416100 - IQMoisture Spare Part Kit Halogen Lamp
  • Metso Slurry Pump - Back Liner
  • Metso 7088010042 - Bronze Lantern Ring
  • Metso IMC-101-M-SC - Industrial Media Converter
  • Voith S1-S1-CAM-ENCL - Metso Pump Cmare Housing & Bracket
  • Metso A4430726V1.1 - Kajaani Connection Box CI 4-9-10
  • Metso A413016 - NCU2 PLC Board
  • Metso D202213P - MBMT80 Base
  • Metso A413135 - AOU-4 Analog Output Board
  • Metso NP 724A - Pneumatic Valve Positioner
  • Metso A413246 - Automation Controller Module
  • Konaflex VKJ-650 - Disc Brake Set Pair
  • Metso WIN3096535D - Core Chuck
  • Metso D201380 - MBR Base with IBC & IPS
  • Metso D202213L - MBMT80 Base Module
  • Metso 181571 - IOP337 PLC Module
  • Metso MF0588336 - Main Hydraulic Filter
  • Metso 181504 - IOP303 RTD Input Module
  • Metso A413240 - PIC2 Board
  • TRICONEX 2211 Baseplate for 3211 I/O Module Triconex System Component
  • TRICONEX CM3201 Communication Module Industrial Control System
  • ABB 560CMR01 Communication Module RTU560 Control Unit
  • Metso 1NWH155 - Ethernet Hub
  • Metso T22PO 103913968 - Sampler Valve (LF02C)
  • Metso EDS-516A-MM-SC - Ethernet Device Switch
  • Metso MC4-16-4 - Brass Rotary Oscillator Drive
  • Metso XT03CWTAS6SJHA - 3" Ball Valve Assembly
  • Metso 181226 - Digital Input Module
  • Metso RAU4E62037 - Valmet Posieve Plus Measuring Head
  • Metso VT-223779/2 - Sealing Housing SD42"B
  • Metso J2029 010 10722 - Circuit Board
  • Metso D202038-ACN-PO - Automation Component
  • Metso NE724/S1-CE01 - Electro-pneumatic Valve Positioner
  • Metso M851004 - CPU Module T194-v0
  • Metso Trucker Hat - Finland Golf Ball Cap
  • Metso DN6-50 - Jamesbury 1/4" Valve
  • Metso B1CU6/20L - Pneumatic Actuator
  • Metso B1CU6/25L - Pneumatic Actuator
  • Metso A416040-09 - Power Rating Unit 28 VDC
  • Metso A4300018B - Central Unit Communication Interface
  • Metso ND9106HN/K05 - Intelligent Valve Controller
  • Metso D200537 BOU83 - Binary Output Model 02
  • Metso A413111 TIU61 - Temperature Input Unit Board
  • Jamesbury 730S313600XTZ2 - 1" Flanged SS Ball Valve
  • Metso HG4A12SH0 - Pressure Transmitter with Coupling
  • Metso A413146 BIU 4 - Binary Input Module
  • Metso DDC7940-D201139 - Controller Module
  • Metso Square Flange - Discharge Flange for Slurry Pump
  • Metso A413177 - Automation Board
  • Metso BOU82 D200536 - Binary Output Module
  • Metso D200619 - Personality Module
  • Metso RKN50MT - Actuator Repair Kit
  • Metso IOP335 - Counter Timer Module
  • Valmet IOP341R 181560 - Positioner Module
  • Metso A413150 - Power Module
  • Metso RKN-184XTZ - Actuator Repair Kit
  • Metso Flinger Seal - Slurry Pump Component
  • Metso Smart-Pulp M2 - Flow Transmitter 24v-dc
  • Metso S420154 - Automation Rack
  • Metso MM0317187 - Efector Kit
  • Metso TCU (MCA) - Consistency Transmitter Analyzer
  • Metso K7504-E - Crusher Pull Back Rod And Nut
  • Metso M2DA02AP - Stainless Flanged Ball Valve 2in
  • Metso vpvl350sr4/5bd - ValvPowr Actuator with ASCO 8551 Valve
  • Metso A413115 TCU41 - Consistency Controller Board
  • Metso D100226 BIC12 - Circuit Board
  • Metso PDWB031 - Bearing Assembly
  • Metso D200532 BIU8 - Binary Input Unit
  • Metso QPXC2/K20C - Spring Diaphragm Actuator
  • Metso PN8504 - Efector Pressure Kit
  • Valmet A413331 SPU - Power Supply Module
  • Metso EDS-518A-MM-SC - Ethernet Switch
  • Metso LK865 - Valve Actuator Mounting Kit
  • Metso LK-909 - Valve Actuator Mounting Kit
  • Metso Z6287570990 - Seal Kit 4501622310
  • Kajaani U4300115A - VME CPU Board Central Unit
  • Metso PO Node - Controller Module