3.7.4 Analyzers with version 4.2.1.14 and later
The Identify feature applies only to PGC5000 Generation 2 version 4.2.1.14 or newer. The OVEN LED
flashing feature will only work if the GC is equipped with the DTC2 and version 4 software. The
Identify feature requires an active connection between the PGC5000 Master Controller and the
associated PGC5000 Oven(s) and a list of compatible analyzers with version 4 software.
Press the Network Connect button and the Connect Dialog box will display a list of compatible
analyzers (see Figure 3-13). Choose an analyzer from the list to display its IP address. Optionally, an
IP address may be entered manually.
Press Connect to initiate a remote connection. Press Cancel to close the dialog box.
Figure 3-13: Connect Dialog Box
When a remote connection is successful, the three Status LEDs on each of the associated PGC5000
Smart Ovens will flash green five times in unison. If available, the Analyzer name will appear in the
Connected To box (see Figure 3-14) If the analyzer name is not available, the IP address will show.
Figure 3-14: Connect box, showing analyzer name
When you press Network Disconnect, the Disconnect dialog box appears (see Figure 3-15). Press
Identify and all three Status LEDs of the remotely connected oven will flash green five times in
unison. Press Disconnect to return to the local connection (when using the Local User Interface) or to
disconnect (when using the Remote User Interface). Press Cancel to close the dialog box.
3.7.5 Master controller as an RUI
When addressing a Master Controller as an RUI, the Status tab will not contain the idle stream button.
The Setup tab will only include the Network Adapters and Clock options. When the Network location in
the System Information Bar indicates “Local,” the Master Controller as an RUI is not connected to an
analyzer.
IMPORTANT NOTE: When connected to a Remote analyzer, the messages that are displayed are
for the remote device.

3.8 Access control list
When the analyzer is shipped from the factory, there are no users configured. If the analyzer is added
to an existing network with other PGC5000s or a STAR DMS, the analyzer’s access control list (ACL)
will be populated by another node in the network. If no other PGC5000s or STAR DMS servers exist in
the network, you can add a user if desired. The first user must be added as a Supervisor.
User passwords should be strong passwords. User names and passwords should be limited to English
keyboard characters, as some special characters are not visible on the LUI and RUI.
3.8.1 Log into the PGC5000 Generation 2 analyzer
Access to the analyzer is controlled by creating authorized user names with passwords. To log into the
analyzer through the LUI:
1. Press the Log In button and select a user name from the list (Figure 3-16).
Figure 3-16: Select a user name
2. Enter your password and press Accept.
3. The information bar at the bottom of the screen displays your access level.
4. The Log In button becomes a Log Out button upon login. To log out, press the Log Out button.