K-WANG



Product positioning and basic information
Omron Z4M is a high-precision infrared laser displacement sensor used for high-speed and precise distance detection, with fast debugging and flexible operation capabilities. It belongs to FDA Class IIIb and IEC Class 3B laser products and needs to be used in conjunction with the Z49-SF2 laser safety kit.
Model and core specifications
Model testing center distance measurement range resolution response speed
Z4M-W40 40mm ±10mm 1.5μm/10μm/40μm 0.15ms/2ms/60ms
Z4M-W100 100mm ±40mm 8μm/30μm/150μm 0.7ms/20ms/500ms
Key performance parameters
Resolution: up to 1.5 μ m (Z4M-W40, 60ms range)
Linearity: Z4M-W40 is 1% FS, Z4M-W100 is 1.5% FS
Output: -4~4V analog quantity, output impedance of 100 Ω, adjustable span to ± 5V
Spot diameter: measuring point Z4M-W40 with a maximum diameter of 1mm, Z4M-W100 with a maximum diameter of 1 × 2mm
Temperature characteristics: Sensor 0.02% FS/℃, amplifier 0.03% FS/℃
Function configuration
Sensitivity switching: WHITE/BLACK/AUTO three levels
Response speed: Three levels can be switched, the faster the speed, the lower the resolution
Status indicator: NEAR/FAR range light, green stability light, red abnormal light
Output hold: When the laser is turned off, the analog quantity and indicator light remain unchanged
Laser delay: Light output 3-10 seconds after power on to ensure safety
Electrical and environmental parameters
Power supply: 12-24VDC ± 10%, ripple ≤ 10%
Power consumption: maximum 120mA
Insulation strength: Sensor 1000VAC/1min, amplifier 300VAC/1min
Working environment: temperature 0~50 ℃, humidity 35%~85%, light ≤ 3000lux
Protection level: IP40
Accessories and supporting equipment
Accessories: Z49-C1 extension cable (3m/8m), Z49-SF2 laser safety kit
Supporting instruments: K3NX, K3TS signal processing table
Output board: K31 series (relay/transistor/linear/communication output)

Wiring and Installation
Wiring definitions: brown (power supply), black (linear output), blue (0V), pink (laser off), orange (enable output)
Wiring requirements: It is prohibited to use the same conduit as high-voltage lines, and the length of the bus cable should be ≤ 10m
Anti interference: Multiple installations must maintain a minimum distance, W40≥30mm,W100≥60mm
Safety regulations
Safe distance: about 1m for Z4M-W40, about 2m for Z4M-W100
Usage requirements: Do not look directly at the laser, FDA warning labels must be posted
Security control: supports triple protection of key switch, interlock terminal, and laser shutdown input
Key questions and answers
Question 1: What is the core difference between Z4M-W40 and Z4M-W100?
Answer: The two models have different measurement center distances, ranges, resolutions, response speeds, and spot sizes; Z4M-W40 has a center distance of 40mm, a range of * * ± 10mm * *, and a maximum resolution of 1.5 μ m; Z4M-W100 has a center distance of 100mm, a range of * * ± 40mm * *, and a maximum resolution of 8 μ m, suitable for different measurement distances and accuracy scenarios.
Question 2: What are the situations where the linear output of Z4M sensor is locked at 6~8V?
Answer: There are three situations: ① a delay period of 3-10 seconds after power on; ② The measured object exceeds the measurement range; ③ Enable the output to be turned off.
Question 3: What are the requirements and protective measures for the safe use of Z4M laser?
Answer: Belonging to FDA IIIb/IEC 3B laser products, direct viewing of the beam is prohibited; The safe distance is about 1m for W40 and 2m for W100. The Z49-SF2 safety kit needs to be installed, and the laser can be controlled using a key switch and interlock terminal. The equipment must be labeled with FDA warning labels to prevent mirror reflection from entering the human eye.

KONG JIANG
Add: Jimei North Road, Jimei District, Xiamen, Fujian, China
Tell:+86-15305925923