+086-15305925923

K-WANG

Service expert in industrial control field!
NameDescriptionContent
Adequate Inventory, Timely Service
pursuit of excellence 
Ship control system
Equipment control system
Power monitoring system
Current position:
新闻动态
newS
   
Brand

Tektronix 5440 oscilloscope

From: | Author:Wang | Time :2025-11-18 | 136 visit: | Share:



Tektronix 5440 oscilloscope

Overview of Instrument Core

(1) Hardware structure

Plugin compartment: There are a total of 3 compartments that can accommodate 5000 series plugins (without N suffix plugins supporting display readings), 2 vertical compartments (left and middle) for amplifier plugins, and 1 horizontal compartment (right) for time base plugins. The combination of plugins determines the functionality of the measurement system.

Key controls:

Display related: INTENSITY adjusts the brightness of the trajectory, READOUT INTENS only adjusts the brightness of the reading part (the reading system fails when counterclockwise to the bottom), FOCUS controls for clear display, and BEAM FINDER compresses the trajectory to the scale range.

Power supply and calibration: The power switch (POWER) controls the power on/off, and the internal calibrator (CALIBRATOR LOOP) provides a square wave signal with a voltage of 400mV (± 1%) and a current of 4mA (± 1%), with a frequency of twice the power supply frequency, for calibration and probe compensation.

External interface: EXT INTENSITY INPUT interface receives Z-axis modulation signals (DC coupling,+5V enhanced brightness, -5V reduced brightness, maximum safe input ± 50V, frequency range DC to 2MHz).

(2) Core functions

Display mode:

Alternate mode: The time base plugin display switch pops up and switches between the activated plugin/channel with each CRT scan, suitable for high-frequency signals.

Chop mode: When the time base plugin display switch is pressed, it electronically switches between channels at a rate of 100kHz, suitable for displaying low-frequency signals or single event phenomena.

X-Y operation: Install 5A series amplifiers or 5B series time base plugins with amplifier channels in the horizontal cabin to receive X signals, and 5A series amplifiers in the vertical cabin to receive Y signals, achieving signal comparison display.

Grating display: Both the left vertical cabin and the horizontal cabin are equipped with 5B series time base plugins. The scanning speed of the vertical cabin is slower than that of the horizontal cabin, and brightness modulation is achieved through EXT INTENSITY INPUT, which can simulate TV display.

Reading System: Only supports 5400 series plugins, displaying alphanumeric information in real-time with analog waveforms on CRT. Each plugin's reading occupies a fixed position (channel 1 is indexed at the top, channel 2 is indexed at the bottom), displaying up to 6 readings (option 3 can select 8 readings). The reading interruption waveform display is only about 20 microseconds/character.

Operation Guide

(1) First operation process (19 steps in total, with the following key steps)

Plug in preparation: Install 5A series amplifier plugs in the left/middle vertical cabin and 5B series time base plugs in the horizontal cabin.

Power connection: After confirming the operating voltage, turn off the power switch and connect a power source that meets the voltage/frequency requirements.

Initial setup:

Amplifier plugin: Display on, center position, CH1 Volts/Div set as needed, CH1 Variable Volts/Div clockwise to the end (Cal), CH1 Input Coupling set to DC, mode set to CH1.

Time base plugin: Display alternately (button pops up), center position, Main Sec/Div set to 5ms, Variable Seconds/Div clockwise to the end (Cal), Swp Mag turned off (button pops up), Main Trig Level counterclockwise to the end, trigger source selected to the left (or center, if the amplifier is in the middle compartment), coupled with Auto Trig, AC Couple,+Slope, mode set to Main Sweep.

Calibration check: Connect the probe/test line to the calibrator, adjust the trigger level to obtain stable display, check the waveform amplitude (4 divisions) and cycle (6 complete cycles, 5 cycles for 50Hz power supply), and recalibrate if abnormal.

(2) Basic measurement methods

Calculation formula/explanation for key steps of measurement type

AC peak to peak voltage: 1. The vertical plug-in input coupling is set to Gnd, and AC is set after receiving the signal; 2. Volts/Div is set to display 5-6 vertical divisions, and Variable Volts/Div is set to Cal; 3. Adjust the trigger to obtain stable display, and measure the vertical division of peak values; 4. Result=Division value × Volts/Div × Probe attenuation ratio (if any) Example: 4.6 Division value × 5V/Div=23V

DC instantaneous voltage: 1. The vertical plug-in input coupling is set to Gnd, and the trajectory is positioned to the reference line (such as the bottom part degree, negative voltage is used at the top); 2. After receiving the signal, set the DC and measure the vertical division between the reference line and the measurement point; 3. Result=division value x Volts/Div x probe attenuation ratio (if any), polarity determined by waveform position (+input and waveform is positive on the reference line) Example: 4.6 division x 2V/Div=+9.2V

Time cycle 1. Connect the signal, select AC/DC coupling, set Volts/Div to display 4 divisions; 2. Adjust the trigger to obtain stable display, set Sec/Div to 1 cycle display<8 divisions; 3. Measure the horizontal division of two points (Variable Sec/Div at Cal); 4. Result=Division value x Sec/Div Example: 5 divisions x 0.1ms/Div=0.5ms

Frequency 1. Obtain the period (T) by measuring the time period method; 2. Frequency=1/T Example: Period 0.5ms, Frequency=1/0.5ms=2kHz

Rising time (10% -90% point) 1. Connect the signal, set Volts/Div and Variable to display 5 vertical divisions; 2. Adjust the trigger to obtain stable display, set Sec/Div to 10% -90% point display<8 divisions; 3. Measure the horizontal division of two points; 4. Result=Division value x Sec/Div Example: 4 divisions x 1 μ s/Div=4 μ s

Specification parameters

1: Vertical amplifier specifications

Additional information on characteristic performance requirements

Bandwidth (6-degree reference) DC to at least 85MHz (067-0680-00 calibration fixture); DC to at least 50MHz (calibrated 5A48)-

Rise time (6 divisions reference) 7ns or less (calibrated 5A48)-

Distortion (6-degree reference) 5% or less (067-0680-00 calibration fixture); 4% or less (calibrated 5A48)-

The influence of position on distortion ± step response signal deviation does not exceed ± 5% (trajectory deviation from the center of the scale does not exceed 1 division)-

Within ± 0.5 divisions of the center of the vertical center scale-

Delay line length 140ns-

Chopping rate 50kHz (+50% -30%), conducting in 3 μ s and turning off in 2 μ s-

Alternating rate: scan every two times-

Table 2: Specifications of Horizontal Amplifier and Z-axis Amplifier

Performance requirements for amplifier type characteristics

Horizontal amplifier bandwidth DC to at least 2MHz (8-division signal reference); At least 20kHz at the low frequency end

The Z-axis amplifier inputs a voltage of+5V to switch the CRT beam from off to on; -5V switches the beam from on to off

Available frequency range DC to 2MHz

Input impedance (resistance) 10k Ω

Table 3: Power, Environment, and Physical Specifications

Category Characteristics Performance Requirements

Power input voltage (RMS) nominal 100V, 110V, 120V, 200V

Input frequency adapted to standard power frequency

Maximum 100W at 120V AC and 60Hz input power

Fuse 120V AC with 1.25A slow melting; 240V AC with 0.7A slow melting

Environmental working temperature: 0 ° C to+50 ° C

Storage temperature -40 ° C to+70 ° C

Working altitude up to 15000 feet

Storage altitude up to 50000 feet

Physical weight 25lbs (11kg)

Panel and shell anodized aluminum panel (gray vinyl coated frame), blue vinyl coated shell

Scale dial 8 × 10 divisions, 1.22cm/division, standard white internal scale line, optional black


Installation and maintenance

(1) Rack installation

Requirement: Suitable for 19 inch standard racks (Universal, EIA, RETMA, Western Electric hole positions), with a minimum vertical space of 5-1/4 inches and an additional 1/4 inch heat dissipation space above and below; Front rail spacing ≥ 17-5/8 inches (unthreaded) or 17-3/4 inches (tapped); The gap between the rear and the rack enclosure is ≥ 2 inches.

Sliding rail installation: The sliding rail is divided into left and right components (marked LH/RH), with the fixed section installed on the front and rear rails of the rack, the chassis section pre installed with instruments, and the middle section connecting the two; During installation, it is necessary to align horizontally to avoid exchanging left and right (which may cause the safety lock to fail and cause the instrument to slide). The front panel needs to be fixed with 10-32 screws (the rack needs to correspond to the hole position).

Maintenance: The slide rail does not require lubrication, and the dark coating on the sliding parts is permanently lubricated.

(2) Plugin installation and maintenance

Installation: Align the upper and lower slots of the plug-in with the cabin guide rail, and push it into the lock; It can be plugged and unplugged with power, without filling all cabins, only installing the required measuring plugs is needed.

Calibration: After the display unit is calibrated according to the instructions, the vertical/horizontal gain is standardized. The plug-in can be interchanged between cabins without the need for readjustment, but the basic calibration of the plug-in itself needs to be checked to ensure measurement accuracy (refer to the plug-in manual).

(3) Safety regulations

Maintenance safety: Only qualified personnel are allowed to carry out repairs, and individual repairs are prohibited (emergency personnel must be provided); Live maintenance should avoid touching exposed connections/components, and disconnect the power supply before dismantling panels, welding, or replacing components.

Power safety: The power supply must meet the RMS voltage between phase lines and between phase lines and ground ≤ 250V, and the grounding conductor of the power line must be reliably grounded.


Performance check

(1) Testing equipment requirements (key equipment as follows)

Equipment Name Performance Requirements Application Scenarios

DC voltmeter range 0-200V (accuracy 0.1%); 0-3000V (accuracy 1%) low/high voltage power supply inspection and adjustment

Calibration generator amplitude calibration 10mV-1V (accuracy ± 0.25%, 1M Ω load), square wave output gain inspection and adjustment

Timing generator outputs 5ns and 10ns markers (accuracy ± 1%) for scanning timing inspection (5ns, 10ns)

5A48 amplifier plug-in bandwidth DC to 500MHz, display mode CH1 vertical/horizontal system adjustment

5B42 time base plugin scanning speed of at least 10ns/div scanning timing inspection and adjustment

(2) Key inspection items

Trajectory alignment: Horizontal trajectories are aligned with the center horizontal scale line, and vertical trajectories are aligned with the center vertical scale line, with an error of ≤ 0.1 divisions.

Geometric shape: Trajectory curvature/inclination ≤ 0.1 divisions, angle between vertical and horizontal trajectories 90 °± 0.7 °.

Vertical performance:

Gain: Input 1kHz square wave, display 5V reference signal, vertical deflection of 5 divisions ± 0.15 divisions (± 3%).

Bandwidth: Input a 50kHz sine wave, display 6 divisions, increase the frequency until the amplitude drops to 4.2 divisions, and the frequency needs to be ≥ 50MHz (5A48 plugin).

Horizontal performance:

Gain: Input 1kHz square wave, display 5V reference signal, horizontal deviation of 5 divisions ± 0.15 divisions.

Bandwidth: Input a 50kHz sine wave, display 6 divisions, increase the frequency until the amplitude drops to 4.2 divisions, and the frequency needs to be ≥ 2MHz.

Reading system: In dual track mode, readings are displayed at the top/bottom of the CRT without overlap, corresponding to Volts/Div and Sec/Div settings; When CAL controls counterclockwise rotation, the ">" symbol is displayed on the left side of the reading.


  • HIMA 895210001 PLC Module | Industrial Safety Control and Processing Module
  • HIMA 157528-0 PLC Module | Industrial Safety Control and Processing Module
  • HIMA B5322 PLC Module | Industrial Safety Control and Processing Module
  • HIMA BV7032 PLC Module | Industrial Safety Control and Processing Module
  • HIMA F3300 PLC Module | Industrial Safety Control and Processing Module
  • HIMA F7105A PLC Module | Industrial Safety Control and Processing Module
  • HIMA F7150 PLC Module | Industrial Safety Control and Processing Module
  • HIMA F7508 PLC Module | Industrial Safety Control Processing Module
  • HIMA F8627 984862765 PLC Module | Industrial Safety Control Module
  • HIMA M3410 PLC Module | Industrial Safety Control Processing Module
  • HIMA Z7308 PLC Module | Safety Control System Processing Module
  • F8640 HIMA Central Module | Safety System Core Controller
  • HIMA F8630 Module | Safety System Functional Module
  • F8651X HIMA Central Module | Safety System Core Controller
  • F3236 Digital Input Module | HIMA Safety System Input Interface
  • F7131 HIMA Power Monitoring Module | Safety System Power Supervision
  • F8627X HIMA Communication Module | Safety System Network Interface
  • HIMA F8560X PLC Module | Safety-Related Control System Component
  • HIMA P8403 PLC Module – Industrial Safety Control Processor
  • HIMA F8628X Communication Module – Industrial Safety System Interface
  • HIMA F8621A Communication Module – Safety System Communication Interface
  • HIMA 984865066 PLC Module – Safety Control Processing Unit
  • HIMA F60CPU01 PLC Module – Safety Control Processing Unit
  • HIMA F8650E PLC Module – Safety Control Processing Unit
  • HIMA F8652E Central Module – Safety Control Processing Unit
  • HIMA F8650X Card – Safety Central Processing Module
  • F8652X HIMA Central Module – Safety Control Central Processing Unit
  • HIMA 99-7105233 B5233-1 Safety Module – Industrial Safety Control Component
  • KEBA KeTop T70-rqa-AK0-LK Teaching Device – Industrial Robot Teach Pendant
  • KEBA KETOP T50-T41-CPU 2495D-0 – Industrial CPU Module
  • KEBA E-CON-CC100/A/ 22178 – Industrial Controller Module
  • KEBA Kemro K2-200 CP 250/X 71580 CPU-MODULE – Industrial Control CPU Module
  • KEBA D3‑DA 330/A‑1211‑00 Drive – KeDrive D3 Axis Controller Industrial Servo Drive
  • KEBA KETOP O70‑bra‑A0a‑F Demonstrator – Industrial Automation Control Module
  • KEBA KETOP O70-BRA-A0A-F Demonstrator | Industrial HMI Operating Terminal
  • KEBA SO84.012.0083.0101.2 Servo Drive | Industrial Motion Control Solution
  • HIMA X-AO1601 Termination Boards
  • HIMA X-AI3251 Digital Output Module
  • HIMA X-DO3251 Digital Output Module
  • HIMA X-DI3202 Termination Board
  • HIMA X-DI6451 Digital Input Module
  • XYCOM 96574-001 - Circuit Board Card Rev A
  • XYCOM 99212A-001 - Control Board Card CPX-7
  • XYCOM 99222-001 - Circuit Board Card Rev A
  • XYCOM 97780-002 - Circuit Board for Operator Panel Screen
  • XYCOM XVME-164/1 - Circuit Board 61116
  • XYCOM CMX-7D - Power Supply Circuit Board PN99865-001
  • XYCOM 10330-00800 - Digital I/O Board Circuit Card
  • XYCOM 8450-HU - Husky Monitor 98916-001
  • XYCOM XVME-428/2 - IASCM Communication Board
  • XYCOM 3512KPT - Industrial PC Operator Interface
  • XYCOM Checkpoint Cognex - Inspection Machine Vision Camera
  • XYCOM XVME-080 - IPROTO Intelligent Prototyping Module
  • XYCOM IV-1653 - Ironics VMEbus CRT Circuit Board
  • XYCOM M032000220 - Control Board
  • XYCOM 4850A - Operator Interface Panel 91855-001
  • XYCOM 9462 - HMI Operator Interface Panel 9462-016214001
  • XYCOM 9465 KPM - Monitor Industrial PC 9465-219114103
  • XYCOM 9486 - Monitor-Miniflex Portrait 9486-0343
  • XYCOM MVME-490/1 - Circuit Board Tegal 6550 Etcher
  • XYCOM 1300 - Node Module 1300-000100000
  • XYCOM 3510 T - Operator Interface
  • XYCOM 3512 KPM - Operator Interface Part No 3512-A1F114103
  • XYCOM 4615KPM - Operator Interface
  • XYCOM 8320 - Operator Interface Display Panel 94321-002
  • XYCOM 4105 - Operator Interface P/N 91904-001
  • XYCOM PM101683E - Operator Interface PM101683 E
  • XYCOM 301993 - Operator Panel 89086-501 Raycon
  • XYCOM 2000 - Operator Panel 97957-001
  • XYCOM 9410KP - Operator Touchscreen HMI 51338-STN
  • XYCOM 9987 - Operator Workstation CPU Board
  • XYCOM 94144-002 - CPU Board Rev V
  • XYCOM 9487 - Programmable Interface HMI Panel PC
  • XYCOM 70956-411 - PLC Module Card Ethernet Card MESA 4I29X
  • XYCOM PM101587 - Operator Panel
  • XYCOM XT1502-BB-RB - Display Panel XT 1502
  • XYCOM PM3510 - Operator Interface Terminal
  • XYCOM PM8450 - Operator Interface Repair Evaluation
  • XYCOM PM8480 - Operator Interface PM101269
  • XYCOM 8000-SKM - Power Supply Board Module
  • XYCOM 510084 - Power Supply Board
  • XYCOM 1546 - Proface Industrial Workstation 1546-102031013
  • XYCOM 86864-002-E - Processor Board Module
  • XYCOM 3112T - Pro-face Operator Interface
  • XYCOM 3712 KPM - Pro-face Operator Panel
  • XYCOM 4615KPMT - Proface Industrial PC Display Panel PM-070007
  • XYCOM 5015T - Pro-Face Touch Monitor 100-240 VAC
  • XYCOM 5015T/R2 - Industrial Flat Panel Touch Monitor 5015R2-0100000
  • XYCOM 1341 - Proface Embedded Computer TF-AEC-6920-C2
  • XYCOM GLC150-BG41-DN - Graphic Logic Controller M-24V
  • XYCOM GP2301H-SC41-24V - Pro-face Operator Interface 5.7" w/ Cable
  • XYCOM 1547 - Pro-face Industrial PC Model 1547-00113101
  • XYCOM ST401-AG41-24V - Graphic Touchscreen Operator Interface
  • XYCOM 3310T - Pro-face Operator Panel 3310-001101001
  • XYCOM 5015 KPMT - Proface Operator Panel Display
  • XYCOM XVME-684 - PC Module VME Bus Processor 70684-201
  • XYCOM 3300 MT - Screen Monitor Display
  • XYCOM SXT1811T - Touch Panel Monitor 100/240VAC
  • XYCOM XT-1502 - Touch Screen Monitor Flat Panel Display Interface
  • XYCOM SXT1811 - Viewtronix SXGA Flat Panel Monitor 18.1" Display
  • XYCOM VT1040T - ViewTronix Touch Screen Monitor
  • XYCOM WS2786291 - Industrial Workstation PC
  • XYCOM XCME-540 - Analog I/O Module VMEbus 70540-001
  • XYCOM XVME-682 - CPU PCB Processor Module 70682-101
  • XYCOM CU040103500 - VME Rack System Module 4200401
  • XYCOM XVME-085 - Proto VMEbus Non-Intelligent Prototyping Module
  • XYCOM XVME-100 - PCB Controller Card ROM Board 70100-001
  • XYCOM XVME-103 - VMEbus Board 10103-001
  • XYCOM XVME-110 - VMEbus Memory Module SRAM 70110-001
  • XYCOM XVME-200 - Digital Universal Instruments I/O Module Card
  • XYCOM XVME-201 - Digital I/O Controller PC Board Card 70201-001
  • XYCOM XVME-202 - VMEbus Control Module 70202-001
  • XYCOM XVME-203 - VME Digital Counter I/O Module Board Card PLC
  • XYCOM XVME-203/293 - VMEbus Counter Module w/ Quadrature
  • XYCOM XVME-212 - 32 Channel Digital Input Programmable Scanner Module
  • XYCOM XVME-220 - 32 Point Programmable Scanner Digital Output Module
  • XYCOM XVME-230 - Intelligent Counter Module Bus Board VMS 70230-001
  • XYCOM XVME-240 - Digital I/O Card VMEbus I/O PCB
  • XYCOM XVME-244 - 64-Channel Isolated Digital I/O Module VMEbus Adept
  • XYCOM XVME-301 - Graphics Controller Module 74301-001A
  • XYCOM XVME-400 - Digital I/O Card VME Module 70400-001
  • XYCOM XVME-401 - VMEbus Circuit Board 70401-001
  • XYCOM XVME-404 - VME Disc Drive Controller Module Board Card PLC
  • XYCOM XVME-428/1 - VME Bus Intelligent Asynchronous Serial Communication Module
  • XYCOM XVME-490/1 - VMEbus Card 71490A
  • XYCOM XVME-491 - VMEbus Module 71491A
  • XYCOM XVME-500 - XVME-500/01 Card 70500-001
  • XYCOM XVME-500/1 - VMEbus Card I/O Module